# Image Artefacts

> Image artefacts in Radiology and Imaging Technology across CT, MRI, ultrasound and nuclear medicine, with causes, geometric clues and corrections for each.

- Canonical URL: https://prepelephant.com/topics/allied/radiology-and-imaging-technology/image-artifacts-rit
- Exam / course: Allied Health · Subject: Radiology and Imaging Technology
- Publisher: PrepElephant (https://prepelephant.com) — Prepared and reviewed by the PrepElephant Academic Review Team
- First published: 2026-10-02
- Last updated: 2026-10-08
- Reference basis: Radiologic Science for Technologists; Christensen's Physics of Diagnostic Radiology
- How to cite: "Image Artefacts", PrepElephant, https://prepelephant.com/topics/allied/radiology-and-imaging-technology/image-artifacts-rit

## Direct answer

Image artefacts are features on an image that do not correspond to real anatomy or pathology, and each modality has its own signature list. CT shows beam-hardening streaks, ring artefacts from miscalibrated detector channels and metal streaks; MRI shows susceptibility voids, chemical shift and phase wrap-around. Ultrasound shows reverberation and acoustic shadowing. Examiners test aetiology, since each cause needs a different correction.

## What you must remember

- **Classification habit:** physics (beam hardening, susceptibility), patient (motion, metal, jewellery), equipment (dead detector channels, zipper lines) and processing (reconstruction streaks, DSA misregistration).
- **CT essentials:** beam hardening causes cupping and dark streaks between dense objects such as the skull base and shoulders; a ring artefact is one faulty detector channel circling the rotation centre; metal streaks come from extreme attenuation plus scatter; photon starvation gives noisy streaks through thick parts.
- **MRI essentials:** susceptibility blooming near metal, air and haemosiderin; chemical shift is water-fat misregistration along the frequency-encoding axis (about 3-4 pixels at standard bandwidth, 1.5 T), with dark-light banding at organ edges; wrap-around is cured by oversampling; motion ghosting runs along the phase axis, Gibbs truncation gives rings at sharp edges, and zipper comes from radiofrequency leakage.
- **Ultrasound essentials:** reverberation gives equidistant repeats in the bladder or a needle; comet-tail comes from crystals and small calculi; mirror image appears above the diaphragm; acoustic shadowing sits behind calculi and gas; posterior enhancement sits behind fluid.
- **Nuclear medicine essentials:** breast, diaphragm and arm attenuation create cold zones; septal penetration gives halos with high-energy isotopes in low-energy collimators; motion causes blur; a hot bladder obscures the pelvis.
- **Radiography and CR/DR:** grid cutoff (upside-down, lateral decentering, off-level) appears as gradual density loss; dead detector lines appear as straight stripes; CR ghosting results from inadequate erasure.
- **Correction logic:** recalibrate for rings; use metal-artefact-reduction algorithms and high-keV images for implants; use breath-hold, gating or faster sequences for motion; oversample for wrap-around.

## Diagnosing an artefact by its geometry

Geometry betrays the origin faster than appearance, so read the pattern before naming the artefact.

1. **Ring on the rotation axis:** a dark ring centred on the gantry's rotation axis, at the same clock position on every slice and in every patient, comes from a detector channel, and recalibration fixes it rather than repeat scanning.
2. **Streaks between dense structures:** dark streaks joining both hips or crossing the skull base in a band are beam hardening, improved by correction kernels.
3. **Straight line across every exposure:** a straight black line running the length of every DR image, regardless of patient, is a dead detector row.
4. **Duplicated anatomy on MR:** a duplicated hand on the opposite side of the image is phase wrap, because the part exceeded the field of view along the phase axis, and oversampling cures it.
5. **Patient contribution:** a pacemaker that shreds a chest MR is susceptibility, invisible on a radiograph, while a hot bladder on pelvic PET is excretion, managed by voiding before scanning.

Physics artefacts yield to algorithms, patient artefacts to preparation, and equipment artefacts to engineering.

## Where students slip

Three errors recur in examinations on this topic.

- **Trap:** describing chemical shift as "fat looking weird". **Correct:** it is misregistration along the frequency-encoding direction with dark-light banding at water-fat interfaces, because fat and water protons differ by about 220 Hz at 1.5 T.
- **Trap:** calling posterior acoustic enhancement an artefact to be removed. **Correct:** it is diagnostically valuable, since fluid transmits sound with less attenuation, so cysts brighten behind; name increased through-transmission as the sign.
- **Trap:** assuming a ring on a gamma camera with a good daily flood is a uniformity fault. **Correct:** a ring that persists despite a perfect flood points to a centre-of-rotation error, which is a geometry problem.

## Frequently asked questions

### What produces a ring artefact on CT images?

A faulty or miscalibrated detector channel produces a ring artefact on CT images. It projects as a circle centred on the rotation axis, persists across slices and patients, and is corrected by recalibration.

### Why does chemical shift artefact occur in MRI?

Chemical shift artefact occurs because fat and water protons precess at slightly different frequencies, about 220 Hz apart at 1.5 T. Fat therefore misregisters along the frequency-encoding axis, giving dark-light banding.

### What is acoustic shadowing on ultrasound, and when is it useful?

Acoustic shadowing is reduced echo return deep to strongly attenuating structures such as calculi or gas. It is diagnostically useful as the signature of gallstones, distinguished as clean or dirty shadowing.

### How is phase wrap-around artefact eliminated on MRI?

Phase wrap-around is eliminated by oversampling in the phase-encoding direction, which stops the aliased anatomy folding back onto the opposite side. Swapping the phase and frequency axes is an alternative.

### Which artefact results from breast or diaphragm attenuation in myocardial perfusion imaging?

Breast and diaphragm attenuation produce a spurious photopenic defect in the anterior or inferior wall of myocardial perfusion images. Prone imaging or attenuation correction should precede any call of ischaemia.