Defibrillation Nursing Care

On this page
  1. Direct answer
  2. What you must remember
  3. A witnessed VF arrest, step by step
  4. Where the traps sit
  5. Frequently asked questions
  6. Related topics

Direct answer

Defibrillation in nursing care starts with the rhythm. Ventricular fibrillation and pulseless ventricular tachycardia are shockable and receive an unsynchronised shock at 120-200 joules biphasic (360 joules monophasic), followed at once by two minutes of CPR. Asystole and pulseless electrical activity are not shockable and receive CPR, adrenaline every 3-5 minutes and a search for reversible causes. Pads go anterolateral or anteroposterior, and the team clears the patient before every shock.

What you must remember

  • Shockable versus non-shockable: VF and pulseless VT are shockable; asystole and PEA are not, since shocking asystole wastes the most valuable minutes of the arrest.
  • Energy doses: biphasic 120-200 J per the manufacturer's recommendation, commonly 200 J, with the maximum available used if unknown; monophasic 360 J; later shocks at equal or higher energy.
  • Rhythm of resuscitation: shock, resume compressions immediately, and reassess the rhythm every 2 minutes; peri-shock pauses are kept to seconds, since survival falls with each interruption.
  • Pad placement: anterolateral (right infraclavicular and left mid-axillary) or anteroposterior; press firmly or shave hairy sites, and keep pads at least 8 cm from an implanted pacemaker.
  • Safety ritual: announce the shock, check visually that no one touches the patient or bed, move oxygen away, dry the chest, then press the discharge button.
  • Synchronised cardioversion: for unstable patients with a pulse, such as broad-complex tachycardia, at lower energy with SYNC engaged; never use SYNC in VF.
  • Drugs around the shock: adrenaline 1 mg IV every 3-5 minutes from the second cycle; amiodarone 300 mg IV after the third shock; treat the 4 Hs and 4 Ts alongside.
  • Nursing ownership: the ICU or CCU nurse performs the daily defibrillator check of battery, pads and self-test, and this audit trail is a legal record.

A witnessed VF arrest, step by step

A post-MI patient in the coronary care unit goes unresponsive, and the monitor shows chaotic ventricular fibrillation. The team follows a fixed order, with the rhythm deciding each action.

  1. Confirm and call: the nurse confirms there is no pulse, starts compressions and calls the code.
  2. Charge and prepare: a second responder charges the biphasic defibrillator to 200 joules while the pads are applied.
  3. Announce and clear: the rescuer says "all clear", and no one touches the bed when the shock is delivered.
  4. Resume at once: hands return to the chest within seconds, with two full minutes of CPR before the next rhythm check.
  5. Post-resuscitation care: the patient converts after the second shock, the airway is supported, oxygenation is brought to normal targets, blood pressure is supported and a twelve-lead ECG is taken.

Where the traps sit

The rhythm decides the action, and the traps follow the rhythm.

  • Trap: shocking a flat line Correct: asystole gets CPR and a search for reversible causes, the 4 Hs and 4 Ts, because electricity does nothing for it.
  • Trap: using SYNC mode in ventricular fibrillation Correct: synchronised shocks wait for an R-wave that VF does not have, so the machine never fires.
  • Trap: reanalysing the rhythm straight after a shock Correct: resume CPR for two minutes first, since the post-shock action is immediate compressions.

Frequently asked questions

Which arrest rhythms are shockable?

Ventricular fibrillation and pulseless ventricular tachycardia are shockable. Asystole and pulseless electrical activity are treated with CPR, adrenaline and correction of reversible causes, not with shocks.

What energy is used for biphasic defibrillation?

Use 120 to 200 joules per the device manufacturer's recommendation, commonly 200 J. If the machine's default dose is unknown, use the maximum available dose.

Why must SYNC mode never be used in ventricular fibrillation?

Synchronised shocks wait for an R-wave to discharge, and VF has no organised R-waves. The machine would never fire while the arrest continues, so the shock would be lost.

Where are defibrillation pads placed?

Place one pad right infraclavicular and the other left mid-axillary at the fifth intercostal space, or use the anteroposterior position. Keep pads at least 8 cm from an implanted pacemaker.

What does the nurse do immediately after delivering a shock?

Resume chest compressions at once and continue CPR for two minutes before reassessing rhythm and pulse. Minimising the peri-shock pause is central to survival.

Practise this in the PrepElephant app

Question banks, previous-year questions, mock tests and revision tools — for Defibrillation Nursing Care and Allied Health Nursing. Free to start.

Get the free app WhatsApp