Radiotherapy Technique Basics
On this page
Direct answer
Radiotherapy technique rests on fractionated dose: curative epithelial cancers conventionally receive 1.8-2 Gy per fraction, five days a week, to 60-70 Gy over 30-35 fractions. Treatment comes from a cobalt-60 unit or a linear accelerator, and the ICRU 50 and 62 framework defines the target volumes and organs at risk. Wedges, compensators, multileaf collimators and bolus then shape the dose to conform to the target.
What you must remember
- Fractionation: curative 1.8-2 Gy per fraction to 60-70 Gy; palliative schedules commonly give 8 Gy in one fraction, 20 Gy in 5, or 30 Gy in 10 for bone metastases.
- Depth of dose maximum: about 0.5 cm for cobalt-60, 1.5 cm for 6 MV, 2.5 cm for 10 MV and 3.0 cm for 15 MV; the build-up region underlies skin sparing.
- Machine physics: cobalt-60 (average 1.25 MeV gammas, 5.26-year half-life) loses output about 1% monthly and is replaced under AERB clearance, while a linac steers an electron beam onto a target for 4-25 MV X-rays and electrons.
- Energy logic: 6 MV suits head and neck, breast and lung; deep pelvic targets gain from 10-15 MV's entrance sparing; electrons treat skin and nodes, with range in centimetres roughly half the energy in MeV.
- ICRU volumes: GTV is visible disease, CTV adds presumed microscopic spread and PTV adds setup and organ-motion margin; organs at risk carry constraint doses that drive plan acceptance.
- Beam modifiers: wedge angle is defined by the isodose tilt at 10 cm depth and compensates for sloping surfaces; bolus of 1-1.5 cm restores surface dose; MLCs shape fields for IMRT and VMAT.
- Technique ladder: 2D planning on bony landmarks, then 3D conformal, intensity-modulated, volumetric arc and image-guided adaptive radiotherapy.
Designing a pelvic treatment plan
A straightforward 3D-conformal prostate plan shows how these choices fit together. Each step narrows the gap between the planned dose and the dose delivered.
- Simulate and contour: The CT scan captures the patient in the immobilisation device, and the oncologist contours prostate and seminal vesicles as GTV, then expands to CTV and PTV and delineates bladder, rectum and femoral heads.
- Choose the beam arrangement: An anteroposterior pair drowns the rectum in mid-dose, so the plan moves to a four-field box (AP, PA and laterals), spreading entrance dose and bringing the rectal V70 into acceptable range.
- Shape the corners: The multileaf collimator shapes the corners of each field around the target.
- Normalise the dose: Normalisation fixes 95% of the prescription dose on the PTV.
- Reproduce the setup: The therapist uses the same immobilisation every day, matches tattoos to room lasers and sets the isocentre at planned coordinates.
- Verify each session: Monitor units are checked independently, and weekly portal or onboard imaging confirms the geometry.
- Run machine QA: Daily output constancy, periodic calibration traceable to national standards and AERB-mandated interlock tests guard against a systematic output error that would multiply across thirty fractions.
Where the exam tries to catch you
Examiners return to the same energy, wedge and machine questions, and each has a classic answer.
- Trap: Saying higher beam energy "penetrates better" as the whole reason for choosing it. Correct: Entrance sparing is the reason deep pelvic targets earn 10-15 MV, while breast treatments stay at 6 MV.
- Trap: Defining wedge angle by the physical filter. Correct: The angle is defined by the isodose tilt at 10 cm depth, which is the answer examiners still expect.
- Trap: Assuming cobalt is obsolete because linacs exist. Correct: Cobalt persists in Indian practice for its simplicity, reliability and lower cost, and its penumbra matters less for bulky disease.
Frequently asked questions
What fractionation is used for curative epithelial tumours?
Curative schedules conventionally give 1.8-2 Gy per fraction, five fractions weekly, to a total of 60-70 Gy in 30-35 fractions. The scheme exploits the differential repair and reoxygenation of normal and malignant tissue.
At what depth does dose maximum occur for cobalt-60 and 6 MV X-rays?
Dose maximum occurs about 0.5 cm deep for cobalt-60 and about 1.5 cm for 6 MV photons. This build-up depth is the basis of skin sparing, which bolus removes when it brings the target to the surface.
How do GTV, CTV and PTV differ?
The GTV is the visible disease, the CTV adds presumed microscopic extension around it, and the PTV adds margins for setup variability and organ motion. Organs at risk are contoured separately and carry their own constraint doses.
Why is a wedge filter used in radiotherapy beams?
A wedge tilts the isodose distribution to compensate for sloping surfaces and to manage hot spots. Its angle is defined by the isodose tilt at 10 cm depth, not by the angle of the physical filter.
Why does a cobalt-60 source need periodic replacement?
Its 5.26-year half-life reduces output by roughly 1% a month, which lengthens treatment times. The source is replaced under AERB clearance rather than run until treatment becomes impractical.