Direct answer

Optical coherence tomography, or OCT, produces cross-sectional images of the retina by measuring the reflection of near-infrared light from its layers. It shows structural change such as macular oedema, macular holes, epiretinal membranes and nerve fibre thinning that the clinical examination may not reveal clearly. Its findings must be read with the clinical picture, because OCT measures structure rather than function.

What you must remember

  • Principle: low-coherence interferometry gives cross-sectional images, with an axial resolution of about 5 micrometres in spectral-domain devices.
  • Normal macula: the foveal depression is present, and the central subfield is typically about 250 to 300 micrometres, depending on the device.
  • Macular oedema: intraretinal cystoid spaces and subretinal fluid, with a thickened central subfield in diabetic and pseudophakic oedema.
  • Macular hole: a full-thickness foveal defect, graded by Gass stage from impending foveal detachment to a full-thickness hole with posterior vitreous detachment.
  • Epiretinal membrane: a hyperreflective band on the inner retinal surface, with a wrinkled, thickened retina and traction on the fovea.
  • Central serous chorioretinopathy: dome-shaped, hyporeflective subretinal fluid, often with a focal defect in the retinal pigment epithelium.
  • Glaucoma: thinning of the peripapillary retinal nerve fibre layer and the ganglion cell layer, which may precede a visual field defect.

Classic OCT findings by condition

OCT images the retinal layers in cross-section, and each disease leaves a characteristic tracing. The table lists the findings that most often appear in exam questions.

Condition Key OCT finding Clinical point
Diabetic macular oedema Intraretinal cystic spaces and a thickened centre Treat when the oedema threatens central vision
Macular hole Full-thickness foveal defect, graded by Gass stage Surgery is considered once the hole is confirmed on OCT
Epiretinal membrane Hyperreflective surface band with a wrinkled retina Symptoms and distortion guide surgery, not the OCT alone
Central serous chorioretinopathy Dome-shaped hyporeflective subretinal fluid Often self-limiting, so observation is common in early cases
Glaucoma Thinning of nerve fibre and ganglion cell layers Structural change can precede visual field loss

Each finding is read against the clinical picture, because a tracing alone does not give a diagnosis. Scans with poor signal strength or segmentation errors can mimic thinning or fluid, so image quality is checked first.

Where the exam tries to catch you

Most errors come from calling a thickness value diagnostic on its own and from confusing a pseudohole with a true hole.

  • Trap: treating a macular hole without checking for an epiretinal membrane Correct: an epiretinal membrane can produce a pseudohole, a hole-like steep foveal contour with the retinal tissue intact and no full-thickness defect, and the surgical plan differs.
  • Trap: treating a single thickness value as diagnostic Correct: values depend on the device and scan pattern, so the number must be read with the scan and the clinical picture.
  • Trap: taking nerve fibre thinning as proof of glaucoma alone Correct: thinning can also follow other optic neuropathies, so it is correlated with the disc and the visual field before a diagnosis is made.

Frequently asked questions

Which OCT finding is typical of central serous chorioretinopathy?

Central serous chorioretinopathy shows dome-shaped, hyporeflective subretinal fluid, often with a focal defect in the retinal pigment epithelium. In most acute cases the fluid resolves over months, so observation is common before treatment is considered.

How does OCT grade a macular hole?

Macular holes are graded by Gass stages, from an impending hole with foveal detachment through to a full-thickness hole with posterior vitreous detachment. The stage helps decide when vitrectomy should be considered, since early stages may close spontaneously.

What is a pseudohole on OCT?

A pseudohole is a hole-like foveal appearance caused by an epiretinal membrane contracting around the fovea, with the retinal tissue intact and no true full-thickness defect. It can look like a macular hole, so the surface membrane must be recognised before surgery is planned.

Which OCT measurements are used to detect early glaucoma?

Peripapillary retinal nerve fibre layer thickness and ganglion cell layer thickness are measured to detect early glaucomatous loss. Thinning can precede a visual field defect, but it must be interpreted together with the optic disc and the field.

Why is the same OCT device used for follow-up?

Retinal thickness values differ between devices and scan patterns, so comparing scans from different machines can mislead. Using the same device and protocol makes thickness changes over time reliable.